Performance Demonstrations of Zeus
In the figures below, we present measurement outcomes obtained through the Zeus Electrochemical Impedance Analyzer with FR-EX3, Enhanced Limitation and High Precision options, All measurements are done with 50cm BNC Cables between sample holder and analyzer. Sampling time 0.5s or 1 Cycle.
Short circuit calibration standard – measured impedance <1mΩ in Basic Range
100Ω calibration standard – measured phase ~10m° in Basic Range
Open electrodes – measured capacity is around 5fF in Basic Range
100pF calibration standard – measured Tan δ < 1e-4 in Basic Range – Correctly compensated in high frequency
1nF Ceramic Capacitor – overcompensated with a shorter leads than short standard
330µH inductor – the model calculates to about 15pF of parallel capacitance inline
10 kΩ SMD Resistor
5pF SMD MLCC
10nF SMD MLCC – High frequency upper measurable capacitor limit hit
Resistor Capacitor network (100pF || [470pF + 20kΩ] || [2nF + 10MΩ])