Performance Demonstrations of Omega – HV6000 Interface

In the figures below, we present measurement outcomes obtained through Omega Advanced Modular Frequency Analyzer with HV6000 High Voltage Impedance Dielectric Interface 2 Wire Front Connection.

All measurements are done with: 
• 100cm BNC Cables between analyzer and Interface
• 50cm SHV Cables between sample holder and analyzer
• Sampling time 0.5s or 1 Cycle.
• Auto Referencing mode always enabled

Omega – HV6000 – short circuit calibration standard – Measured impedance < 20 Ω

Omega – HV6000 – open SHV connector – measured capacity < ~20fF in Basic Range

Omega – HV6000 – 100Ω calibration standard – measured deviation in impedance < 30Ω

Omega – HV6000 – 1kΩ – measured with 2121Vrms applied voltage

Omega – HV6000 – 100pF Polyester Capacitor – measured Tan δ < 1e-4 in Basic Range

Omega – HV6000 – 47pF Ceramic Capacitor – measured with 2121Vrms applied voltage

Omega – HV6000 – 330pF Ceramic Capacitor – measured with 2121Vrms applied voltage

Omega – HV6000 – 1nF Ceramic Capacitor with 2121Vrms applied voltage

Omega – HV6000 – 470pF Ceramic Capacitor – measured with +1500 Vdc and 1000Vrms applied voltage

Omega – HV6000 – Resistor Capacitor network (100pF || [470pF + 20kΩ] || [2nF + 10MΩ])