Performance Demonstrations of Lamda

In the figures below, we present measurement outcomes obtained through the Lamda specialized Impedance Analyzer with FR-EX3, Enhanced Limitation and High Precision options, All measurements are done with 100cm BNC Cables between sample holder and analyzer. Sampling time 0.5s or 1 Cycle.

Short circuit calibration standard – measured impedance < 10mΩ in Basic Range

100Ω calibration standard – measured deviation < 0.5 Ω

Open Iin BNC connector – measured capacity is around 5fF in Basic Range

100pF calibration standard – measured Tan δ < 1e-4 in Basic Range – Correctly compensated in high frequency

1nF Polyester Capacitor – undercompensated with an inductance of a wire in series

5pF SMD multi-layer ceramic capacitor

0.6pF (2 parallel 1.2pF) multi-layer ceramic capacitor

Resistor Capacitor network (100pF || [470pF + 20kΩ] || [2nF + 10MΩ])